Vertical / Horizontal Flame Apparatus Lab Testing Equipment 5V / HB/ V-0/1/2 UL94

  • 1
  • 2

Brand Name : PEGO GROUP

Model Number : HVR-4H

Certification : Third-lab calibration certificate

Place of Origin : China

MOQ : 1 set

Price : USD3100-USD4500/set

Payment Terms : T/T, Western Union, MoneyGram

Delivery Time : 15 working days

Packaging Details : each packed in a wooden case

Reference Standard : UL94, IEC60695-11-4 and IEC60695-11-3

Production name : flammability test equipment

Application : V-0/1/2, HB, 5VA/B, HF-1/2 and HBF

Volume of chamber : 0.75CMB

Control units : single chip and button control

Dimension : 1100mm(W)*700mm(D)*1300mm (H)

Working power : single phase 220V/50-60Hz

Warranty : 12 months

Contact Now

5V / HB/ V-0/1/2 UL94 Vertical and Horizontal Flame Test Apparatus Lab Testing Equipment

1. Introduction of Flame test Apparatus

Horizontal and vertical flame test apparatus is applied to 50W vertical flame test includes V-0, V-1, V-2, horizontal flame tester HB, and 500W vertical flame test includes 5VA, 5VB, HF-1, HF-2 and HBF. It is widely used in the R&D department, QC department and production department of lighting equipment, low-voltage apparatus, household appliance, motor and etc.. And also can be used in the industries of insulation material, plastic, non-metalic material, foam with a density of not less than 250kg/m' and other solid combustible material.

Horizontal and vertical flame test apparatus is applied to simulate the fire at the early stage surround the electric and electronics appliance, and then evaluate the degree of fire hazard. It is designed according to UL94, IEC60695-11-4 and IEC60695-11-3. The flammability test uses a bunsen burner of a specified size and specified gas source (methane or natural gas) to ignite the horizontal or vertical specimen for several times in a certain flame height and a certain flame angle, the fire hazard is evaluated by burning rate, after glow time, after flame time and damage length of the specimen.

The horizontal and vertical flame test apparatus is controlled by single chip and push button. The test apparatus is built-in air exhaust, the exhaust system can be preset to delay 1-3mins after power off. The front panel is made by military-grade explosion-proof glass and transparent for easy observation of the experimental process, and the shell is made by coated hardness steel. Fully independent electrical circuit and gas circuit design for safer use.

2. Parameter of Flame Test Apparatus

Burner Φ9.5mm ± 0.3mm, 100mm length bunsen burner, source: 50W and 500W
Dip angle 00, 200, 450 (manual adjustment)
Flame height 20mm ± 2mm to 180mm±10mm (adjustable), record by height gauge
Flame time 0-999.9s±0.1s (adjustable)
After flame time 0-999.9s±0.1s (auto record, manual pause)
After glow time 0-999.9s±0.1s (auto record, manual pause)
Number of flame 0~9999 times (presettable)
Fuel gas 98% methane or 37MJ/m3±1MJ/M3 natural gas
Flow and pressure record Double flowmeters and manometer
Gas Pressure 0.1Mpa
Gas Flow 50W:105±5ml/min or 18±1 mm, 500W:965±30ml/mino or 117±3 mm
Flame calibrations

1). Flame temperature required: heat up from 1000C±50C to 7000C±30C within the time of 44s±2s (50W) and 54s±2s (500W)

2). Flame height: 50W: 20±2mm (50W) 500W: 40±2mm (inner), 125±10mm (outer)

3). Thermocouple: Φ0.5mm K type thermocouple (The insulating sleeve heat resist to 1250 ℃), 2 pieces to detect the temperature

4). Temperature measuring copper blocks

Test background black
Volume of chamber 0.75 CBM, exhaust hole: Φ70mm
Thermal radiation baffle Automatic masking
Safe operation interface automatically power off upon open the glass window
Accessories Pick-up tray, sample clamp, wrapping tissue 12g/m2 ~30g/m2, steel wire mesh, height gauge, Cotton and etc.
Dimension of apparatus 1100mm(W)*700mm(D)*1300mm (H)
Power supply AC220V, 50Hz, ≤500W

3. FAQ

1). What is the warranty period of the equipment?
Answer: the equipment warranty is 12 months upon you receipt of the equipment

2). Can you provide third-lab calibration certificate?
Answer: Yes, the equipment can send to the third-labs authorized by ISO17025 and get the calibration certificate, but the buyer should undertake the charges.

3). What is the working power of the equipment?
Answer: single phase 220v/50-60Hz

4). What is the language of the operation interface?

Answer: English


Product Tags:

electrical testing instruments

      

electric test meter

      
Quality Vertical / Horizontal Flame Apparatus Lab Testing Equipment 5V / HB/ V-0/1/2 UL94 for sale

Vertical / Horizontal Flame Apparatus Lab Testing Equipment 5V / HB/ V-0/1/2 UL94 Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: Pego Electronics (Yi Chun) Company Limited
*Subject:
*Message:
Characters Remaining: (0/3000)